Technical specifications |
310A |
410A |
510A |
620M |
120CW |
|
Tissue/Vessel size |
>60 μm |
>60 μm |
>60 μm |
>60 μm / >450 μm up to 10 mm |
>60 μm |
|
Chamber(s) |
Single bath |
Dual, independent or common |
Dual, independent or common |
Four individual chambers |
Single bath/conical shape |
|
Chamber material |
Acid-resistant stainless steel |
Acid-resistant stainless steel |
Acid-resistant stainless steel |
Acid-resistant stainless steel |
Acid-resistant stainless steel |
|
Chamber volume |
Max. 10 ml |
Max. 10 ml |
Max. 10 ml |
Max. 8 ml |
Max. 10 ml |
|
Chamber suction |
No |
No |
No |
Manual or automatic, time controlled, user defined |
No |
|
Chamber cover |
Yes |
With connection for suction and gassing |
With connection for suction and gassing |
Supplied with connections for gassing |
With connection for suction and gassing |
|
Force range |
-100 -/+200 mN |
-100 - /+200 mN |
-100 - /+200 mN |
User selectable at +/-200/400/800 mN |
-100 - /+200 mN |
|
Force resolution |
0.01mN |
0.01mN |
0.01mN |
0.01mN |
0.01mN |
|
Micropositioners |
Manually operated |
Manually operated |
Automatical linear motor operated |
Manually operated |
Manually operated precision micrometer |
|
Weight calibration |
Manual |
Manual |
Manual |
Semi Automatic |
Manual |
|
Heating |
Built into chamber,
independent of superfusion
|
Built into chamber,
independent of superfusion
|
Built into chamber,
independent of superfusion
|
Built into chamber,
independent of superfusion
|
Built into chamber,
independent of superfusion
|
|
Temp. range |
Ambient temp.
to 50 °C |
Ambient temp.
to 50 °C |
Ambient temp.
to 50 °C |
Ambient temp.
to 50 °C |
Ambient temp.
to 50 °C |
|
Temp. resolution |
0.1 °C |
0.1 °C |
0.1 °C |
0.1 °C |
0.1 °C |
|
Temp. probe |
External |
External |
External |
External |
External |
|
Output reading |
Force (mN) |
Force (mN) |
Force (mN) |
Force (mN) |
Force (mN) |
|
Analog output |
Up to four output, 1.0V full scale for all acquired signals, user defined |
Up to four output, 1.0V full scale for all acquired signals, user defined |
Up to four output, 1.0V full scale for all acquired signals, user defined |
Independent filtered 4 channel output at 1.0V full scale |
Up to four output, 1.0V full scale for all acquired signals, user defined |
|
PC connection |
Serial interface - RS232/RS485 |
Serial interface - RS232/RS485 |
Serial interface - RS232/RS485 |
USB 2.0 |
Serial interface - RS232/RS485 |
|
Voltage |
100 to 240 VAC (auto) 50/60 Hz |
100 to 240 VAC (auto) 50/60 Hz |
100 to 240 VAC (auto) 50/60 Hz |
External power supply |
100 to 240 VAC
(auto) 50/60 Hz |
|
Ambient temp. |
15-30 °C |
15-30 °C |
15-30 °C |
15-30 °C |
15-30 °C |
|
Optional accessories |
|
|
|
|
|
|
- range |
pH 0 - 14 |
pH 0 - 14 |
pH 0 - 14 |
|
pH 0 - 14 |
|
- temp. correction |
0 - 50 °C |
0 - 50 °C |
0 - 50 °C |
|
0 - 50 °C |